All Tru-Test load bars will be automatically detected by the S2 or S3 indicator. Most leading load bars can also be set up to work with the indicator. Select your load bars from the Load bar type option on the Settings page of the Data Link app or Data Link for PC. However, if your load bars are not available for selection, you will need to perform a span calibration to set up your load bars to weigh correctly.


What you need

  • A known weight (e.g. a few bags of feed with accurate, known weights)
  • Ensure that your load bars are free from dirt and stones
  • Know the capacity of your load bars. If you don't know the capacity of your load bars, you can calibrate the indicator by using a weight that is slightly greater than your heaviest animal. Enter this in the capacity field during the span calibration process.

You can perform a span calibration from the Settings page by following the step-by-step instructions.

If there are settings for a prior span calibration you can erase it with the Clear Calibration option.

You can perform a span calibration from the Settings page by following the steps.

If there are settings for a prior span calibration you can erase it with the Clear Span Calibration option.